70811_18.pdf
(
255 KB
)
Pobierz
CHAPTER 18
CALIBRATION OF PICKUPS
M. Roman Serbyn
Jeffrey Dosch
INTRODUCTION
This chapter describes various methods of calibrating shock and vibration transduc-
ers, commonly called
vibration pickups.
The objective of calibrating a transducer is
to determine its sensitivity or calibration factor, as defined below. The chapter is
divided into three major parts which discuss comparison methods of calibration,
absolute methods of calibration, and calibration methods which employ high accel-
eration and shock. Field calibration techniques are described in Chap. 15.
PICKUP SENSITIVITY, CALIBRATION FACTOR,
AND FREQUENCY RESPONSE
As defined in Chap. 12, the
sensitivity
of a vibration pickup is the ratio of electrical
output to mechanical input applied along a specified axis.
1–3
The sensitivity of all
pickups is a function of frequency, containing both amplitude and phase informa-
tion, as illustrated in Fig. 18.1, and therefore is usually a complex quantity. If the sen-
sitivity is practically independent of frequency over a range of frequencies, the value
of its magnitude is referred to as the
calibration factor
for that range, but it is speci-
fied at a discrete frequency. The phase component of the sensitivity function likewise
has a constant value in that range of frequencies, usually equal to zero or 180°, but it
may also be proportional to frequency, as explained in Chap. 12.
The
frequency response
of a pickup is shown by plotting the magnitude and phase
components of its sensitivity versus frequency. This information is usually presented
relative to the value of sensitivity at a reference frequency within the flat range. A
preferred frequency, internationally accepted, is 160 Hz.
Displacements are usually expressed as single-amplitude (peak) or double-
amplitude (peak-to-peak) values, while velocities are usually expressed as peak,
root-mean-square (rms), or average values. Acceleration and force generally are
expressed as peak or rms values. The electrical output of the vibration pickup may
be expressed as peak, rms, or average value. The sensitivity magnitude or calibration
factor are commonly stated in similarly expressed values, i.e., the numerator and
18.1
18.2
CHAPTER EIGHTEEN
0
°
30
PHASE RESPONSE
dB
–30
°
20
–60
°
10
–90
°
0
AMPLITUDE (SENSITIVITY)
RESPONSE
–120
°
–10
–150
°
–20
–180
0.05
–30
0.1 0.2 0.3 0.4 0.5
PROPORTION OF MOUNTED RESONANCE FREQUENCY f
m
1
1.5
2
3
4
5
FIGURE 18.1
Pickup amplitude and phase response as functions of frequency. (
After M. Ser-
ridge and T.R. Licht.
4
)
denominator are both peak or both rms values. Examples of typical sensitivity spec-
ifications for an accelerometer: 2 pC per m/sec
2
, 10 millivolts/m/sec
2
, 5 milliV/
g
(where C is the symbol for coulomb, V is the symbol for volt, and
g
is the accelera-
tion of gravity). For some special applications it may be desirable to express the sen-
sitivity in mixed values, such as rms voltage per peak acceleration.
CALIBRATION TRACEABILITY
In calibrating an instrument, one measures the instrument’s error relative to a refer-
ence which is traceable to the national standard of a country. A calibration is said to be
traceable
5
to a national or international standard if it can be related to the standard
through an unbroken chain of comparisons—all having stated uncertainties. In the
U.S.A., for example, national vibration standards are maintained at the National Insti-
tute of Standards and Technology in Gaithersburg, Maryland. A number of other
national metrology laboratories having known capabilities for maintaining national
vibration standards are listed in Table 18.1. Countries whose national laboratories do
not provide a national vibration standard may belong to a regional international asso-
ciation, such as NORAMET (North American Metrology Cooperation), EUROMET
(European Metrology Cooperation), or OIML (Organization for Legal Metrology)
that can assist transducer manufacturers in setting up steps necessary for establishing
traceability to a national standard.
Vendors of transducers must be able to show that calibrations of their instru-
ments are traceable to a national standard by means of calibration reports stating
the value(s) of sensitivity, measurement uncertainty, environmental conditions, and
18.3
CALIBRATION OF PICKUPS
TABLE 18.1
National Standards Laboratories Responsible for the Calibration
of Vibration Pickups
Institution
Laboratory
Location
Country
CSIRO
Natl. Measurement Laboratory
Lindfield
Australia
INMETRO
Laboratório de Vibrações
Rio de Janeiro
Brazil
NRC-CNRC
Inst. Natl. Meas. Stds.
Ottawa
Canada
NIM
Vibrations Laboratory
Beijing
China
CMU
Primary Stds. of Kinematics
Prague
Czech Repub.
B&K
Danish Prim. Lab. for Acoustics
Naerum
Denmark
BNM
CEA/CESTA
Belin-Beliet
France
PTB
Fachlabor. Beschleunigung
Braunschweig
Germany
IMGC
Sezione Meccanica
Torino
Italy
NRLM
Mechanical Metrology Dept.
Tsukuba
Japan
KSRI
Division of Appl. Metrology
Daedeog Danji
Rep. of Korea
NMC
SIRIM
Berhad
Malaysia
CENAM
Div. Acustica y Vibraciones
Queretaro
Mexico
DSIR
Measurement Stds. Laboratory
Lower Hutt
New Zealand
VNIIM
Mendeleyev Inst. for Metrology
St. Petersburg
Russia
ITRI
Center for Measurement Stds.
Hsinchu
Taiwan
NIST
Manufacturing Metrology Div.
Gaithersburg
U.S.A.
identification of the standard(s) used in the calibration procedure. Depending on
the application, there may be one or more links to the national standard.
Primary and Secondary Standards.
Primary standards,
6,7
maintained at national
metrology institutes, are derived from
absolute measurements
8
of the transducer’s
sensitivity, measured in terms of seven basic units. For example, the absolute meas-
urement of “speed” must be made in terms of measurements of distance or distance
and time, not by a speedometer. Thus the word
absolute
implies nothing about preci-
sion or accuracy. An example of a laboratory setup for the calibration of primary-
standard accelerometers, derived from absolute measurements, is shown in Fig.
18.2.
9
A vibration exciter generates sinusoidal motion which is measured by a
ACCELEROMETER
STANDARD
AIRBORNE
ACCELERATION EXCITER
INTERFEROMETER
LASER
SIGNAL-
PROCESSING
SYSTEM
LIGHT
DETECTOR
DUMMY
MASS
INDICATING
INSTRUMENT
(STANDARD)
FIGURE 18.2
Primary (absolute) calibration of an accelerometer standard using laser interferom-
etry. (
After von Martens.
9
)
18.4
CHAPTER EIGHTEEN
Michelson interferometer (described later in this chapter). The vibration is applied
to the base of the standard accelerometer whose output is measured. A dummy
mass, mounted on its top surface, simulates the conditions when this standard
accelerometer is used to calibrate a
secondary standard
10
accelerometer by the com-
parison method described in the next section.
Secondary standards
(also referred to
as
transfer standards
or
working standards
) are maintained at various government
laboratories and industrial laboratories. A secondary standard accelerometer may be
calibrated either from absolute measurements or from a comparison with a primary
standard accelerometer. Such secondary standards are usually used for purposes of
comparisons of calibrations between laboratories or for checking production and
field units.
COMPARISON METHODS OF CALIBRATION
A rapid and convenient method of measuring the sensitivity of a vibration pickup to
be tested is by direct comparison of the pickup’s electrical output with that of a sec-
ond pickup (used as a “reference” standard) that has been calibrated by one of the
methods described in this chapter. A comparison method is used in most shock and
vibration laboratories, which periodically send their standards to a primary stan-
dards laboratory for recalibration. This procedure should be followed on a yearly
basis in order to establish a history of the accuracy and quality of its reference stan-
dard pickup.
In this method of calibration the two
pickups usually are mounted back-to-
back on a vibration exciter as shown in
Fig. 18.3. It is essential to ensure that
each pickup experiences the same
motion. Any angular rotation of the
table should be small to avoid any dif-
ference in excitation between the two
pickup locations. The error due to rota-
tion may be reduced by carefully locat-
ing the pickups firmly on opposite faces
with the center-of-gravity of the pickups
located at the center of the table. Rela-
tive differences in pickup excitation may
be observed by reversing the pickup
locations and observing if the voltage
ratio is the same for both positions.
Calibration by the comparison method is limited to the range of frequencies and
amplitudes for which the reference standard pickup has been previously calibrated.
If both pickups are linear, the sensitivity of the test pickup can be calculated in both
magnitude and phase from
FIGURE 18.3
Comparison method of calibra-
tion: Pickup 2 is calibrated against Pickup 1 (the
reference standard). The two pickups may be ex-
cited by any of the means described in this chap-
ter. (
After ANSI Standard S2.2-1959, R 1997.
1
)
e
t
e
r
S
t
=
S
r
(18.1)
where
S
t
= sensitivity of test pickup
S
r
= sensitivity of reference standard pickup
e
t
= output voltage from test pickup
e
r
= output voltage from reference standard pickup
18.5
CALIBRATION OF PICKUPS
Several calibration methods described below are variations on the implementation
of Eq. (18.1); they differ mainly in the manner of vibration excitation.
USING THE COMPARISON METHOD
A simple and convenient way of performing a comparison calibration is to fix the
test pickup and reference standard pickup so they experience identical motion, as in
Fig. 18.3. Then, set the frequency of the vibration exciter at a desired value, adjust the
amplitude of vibration of the vibration exciter to a desired value, and then compare
the electrical outputs of the pickups. Often, instead of making a comparison at a
fixed frequency, a graphical plot of the sensitivity versus frequency is obtained by
incorporating a swept-frequency signal generator in the calibration system.
RANDOM-EXCITATION-TRANSFER-FUNCTION METHOD
The use of random-vibration-excitation and transfer-function analysis techniques
can provide quick and accurate comparison calibrations.
11
The reference standard
pickup and the test pickup are mounted back-to-back on a suitable vibration exciter.
Their outputs are usually fed into a spectrum analyzer through a pair of low-pass
(antialiasing) filters. The bandwidth of the random signal which drives the exciter is
determined by settings of the analyzer.
This method provides a nearly continuous calibration over a desired frequency
spectrum, with the resulting sensitivity function having both amplitude and phase
information. Since purely sinusoidal motion is not a requirement as in the other cal-
ibration methods, this lessens the requirements for the power amplifier and exciter
to maintain low values of harmonic distortion. A very useful measure of process
quality is obtained by computing the input-output coherence function, which
requires knowledge of the input and output power spectra, the cross-power spec-
trum, and the transfer function.
CALIBRATION BY ABSOLUTE METHODS
RECIPROCITY METHOD
The reciprocity calibration method is an absolute means for calibrating vibration
exciters that have a velocity coil or reference accelerometer. This method relates the
pickup sensitivity to measurements of voltage ratio, resistance, frequency, and mass.
For this method to be applicable, it is necessary that the vibration exciter system be
linear (e.g., that the displacement, velocity, acceleration, and current in the driver
coil each increase linearly with force and driver-coil voltage). The reciprocity
method is used chiefly with electrodynamic exciters
12
but also with piezoelectric
vibration exciters.
13
The reciprocity method generally is applied only under controlled laboratory
conditions. Many precautions must be taken, and the process is usually time-
consuming. Several variations of the basic approach have been developed at
national standards laboratories.
14,15
The method described here has been used at the
Plik z chomika:
rajmundos9
Inne pliki z tego folderu:
0071481133_ar002.pdf
(1447 KB)
0071481133_ar003.pdf
(1418 KB)
0071481133_ar004.pdf
(1487 KB)
0071481133_ar005.pdf
(1187 KB)
0071481133_ar001.pdf
(355 KB)
Inne foldery tego chomika:
# Kurs języka angielskiego -1000 godzin nauki PL
# Niemiecki
_ Sieci Instalacje Maszyny
02 Robert Kiyosaki - Kwadrant Przepływu Pieniędzy
03 Robert Kiyosaki - Inwestycyjny Poradnik Bogatego Ojca
Zgłoś jeśli
naruszono regulamin